Autosorb®-6B
Specifications & Physical Requirements

|
| Specification |
Autosorb-6B |
PRESSURE (Specifications by
transducer manufacturer ) |
Range: 0 - 1,000 torr
|
| |
Accuracy: |
0.1% full scale |
|
Linearity: |
0.5% full scale |
|
Pressure Resolution: |
0.00026 torr |
|
Minimum P/Po: (N2) |
0.0001 |
ULTIMATE VACUUM: |
5 x10-3 mbar ( two stage rotary pump - included). |
|
ADSORBATES: |
Nitrogen or any non-corrosive gas with appropriate coolant |
SURFACE AREA RANGES: |
0.01 m2/g with N2 to no known upper limit |
PORE SIZE RANGE: |
3.5 to >4000Å
|
MINIMUM PORE VOLUME (Liquid):
(STP):
|
1 x10-6cc/g
5 x10-5cc/g
|
|
| ACCESSORIES: |
|
| Required(not included): |
Adsorbate and helium (minimum purity 99.999%), coolant (eg. LN2), PC (Windows), degasser (Autosorb Degasser, MasterPrep, FloVac or Flow Degasser). |
| Optional: |
Uninterruptable power supply,
20 inches (50cm) high caster mounted cabinet. |
|
| DIMENSIONS: |
|
| Width
|
25.5 inches (64.8 centimeters) |
|
Height
|
45.5 inches (115.6 centimeters) |
| Depth
|
30.0 inches (76.2 centimeters) |
| WEIGHT: |
380 pounds (172.7 kilograms) |
| ELECTRICAL: |
100-240 volts, 50/60 Hertz |
|
| AMBIENT: |
10 - 38°C. operating range |
| |
90% maximum relative humidity |
| |
|
|
Return to Autosorb-6B
|
|
|
 |
 |
|
 |
 |
|
|
|
|
|
© Quantachrome Instruments, 2006-2012. All Rights Reserved. |
|
|