Technologies Instruments Lab Services About Us Directory Contact Us
Quantachrome Instruments: Porous Materials Characterization - Surface Area, Pore Size, Density
Product Listing, click here
Spare Parts Catalog
Science
Technical Resources
What's New
Training Schedule
Events Calendar
Notes and Articles
Join Our Mailing List

QUANTACHROME INSTRUMENTS
1900 Corporate Drive
Boynton Beach, Florida 33426 USA

U.S. and Canada
1-800-989-2476

Worldwide
Tel: 1-561-731-4999
Fax: 1-561-732-9888

 

Print Print this page
Home
 

Autosorb-1 Series Specifications

Questions-Feedback

Pressure Transducers (per manufacturer's specifications)
Autosorb
-1
Autosorb
-1-MP
Autosorb
-1-C
0 to 1,000 torr
yes
yes
yes

Accuracy: 0.1% full scale
Linearity: 0.05% full scale
Minimum resolvable pressure: 0.00025 torr
Minimum resolvable relative pressure (P/Po): 3.2 x 10-7(nitrogen)

0 to 10 torr
no
yes
yes

Accuracy: 0.15% reading
Linearity: 0.15% reading
Minimum resolvable pressure: 2.5 x 10-6 torr
Minimum resolvable relative pressure (P/Po): 3.2 x 10-9(nitrogen)

0 to 1 torr
no
yes
yes

Accuracy: 0.15% reading
Linearity: 0.15% reading
Minimum resolvable pressure: 2.5 x -7 torr
Minimum resolvable relative pressure (P/Po): 3.2 x 10-10(nitrogen)

Ultimate Vacuum (per manufacturer's specifications)

Roughing Pump

<10mtorr
<2mbar
<2mbar

Dry Roughing Pump

no
yes
yes

Oil-Free Turbo Molecular Pump: <3.75 x 10-10 torr
(manufacturer's specification)

no
yes
yes
Adsorbates

Non-corrosive gas (N2, Ar, CO2, C4H10 , etc.)

yes
yes
yes

Krypton gas

no
yes
yes

Ammonia

no
no
yes
Vapors (water and organic solvents)
no
option
option

Programmable 5-gas switching

no
option
yes
Temperature Capabilities
Degas Heating Mantles (2 stations)
yes
yes
yes

Heating up to 400°C in steps of 1°C
Accuracy within 5°C of set point
Optional quartz mantles for higher temperatures

Optional programmable ramp/multi-step controllers

Analysis Heating Furnace (Autosorb -1-C)
no
no
yes

Heating to 1,100°C with automatic calibration
Programmed ramp rates from 1 to 50°C/min.
Unlimited number of heating/cooling programs

Auto-Leveling Cooling Bath
yes
yes
yes

Two liter standard (60 hours)
One liter, wide mouth (optional)

Cryostat
no
option
option
More Autosorb-1 Series Specifications | Return to Autosorb-1 Series
P.R.China S. Korea Japan Germany UK & Ireland Alfatest Elexience Turkey Canada Brasil Hellas (Greece) Scientific Solutions
© Quantachrome Instruments, 2006-2008. All Rights Reserved.
ISO 9001: 2000